Measurement, control and laboratory equipment
IEC 61326-1 EMC for measurement, control and laboratory equipment — including the performance criteria that decide whether a disturbed instrument still reads correctly.
- IEC / EN 61326-1
- GB/T 18268.1
- CISPR 11 / EN 55011
- IEC 61000-4-2 / -4 / -5 / -6 / -8 / -11
- IEC 61000-3-2 / -3-3
One unit in its final enclosure; describe the functions to be monitored during immunity
8–12 working days, depending on the number of monitored functions
What this covers, and why it matters
The short version before the detail — product families, typical use, and the practical difference our scope makes.
- Analytical and laboratory instruments
- Measurement and test equipment
- Sensors, transmitters and controllers
- Data acquisition units
- Industrial monitoring devices
- CE EMC files for instruments
- Immunity verification for process equipment
- Customer acceptance testing
- Field instrument qualification
- Performance criteria fixed before the run
- Function-level results, not just a pass line
- Safety and EMC mapped in one scope check
An instrument that survives an electrical disturbance but reports a wrong value has failed its purpose. That is why the EMC standard for measurement and control equipment is built around performance criteria, not just pass or fail. The criterion has to be agreed before the test starts.
Product coverage
We assess electrical equipment for measurement, control and laboratory use under IEC / EN 61326-1 and the GB/T 18268 equivalent. This covers analytical instruments, sensors and transmitters, data acquisition units, controllers, and bench equipment used in laboratory and industrial settings.
- Emissions to CISPR 11 / EN 55011, with the class matched to the use environment.
- Immunity to the IEC 61000-4 series at the levels defined by IEC 61326-1.
- Functional performance monitored during and after each disturbance.
- Additional part 2 requirements where the instrument has a specific function.
Performance criteria are the point
IEC 61326-1 defines what happens during immunity: continued operation without degradation, temporary loss of function that self-recovers, or loss that requires intervention. We agree the applicable criterion with you for each function, because the answer depends on whether the instrument is monitoring a process or merely displaying a reading.
Where it applies
Laboratory and field instruments, process transmitters, test and measurement devices, and control equipment intended for industrial or light-industrial environments. Products in this family often need both the generic instrumentation standard and a product-specific safety assessment.
What the report tells you
A criterion table agreed before testing, results by phenomenon with the observed behaviour, and a report that states whether any deviation recovered on its own. When a deviation needs intervention, we identify the disturbance that caused it.
What you receive
- Performance-criteria table
- Emissions and immunity report
- Functional behaviour record per phenomenon
- Deviation and recovery notes
Frequently asked
What is a performance criterion?
It defines what the instrument is allowed to do during a disturbance. We agree it per function before the test, so the result has a clear meaning.
Do you test the measurement accuracy as well?
We monitor the agreed functions during immunity. Accuracy calibration is a separate service and can be arranged if required.
Is EN 61326-1 enough for a CE declaration?
Often yes for EMC, but the safety standard for the instrument type is separate. We check both when you enquire.